High Resistance / Low Conductance Meter HRLC

Updated:2025-03-13

Views:2271

  • Brand: ALPHALAB
  • Model: HRLC
  • Description: There is always a problem with ultra high resistance measurements: only a very small amount of current is flowing through the component that is being tested. Therefore certain precautions must be take
  • Feedback
  • WeChat

    QQ

    Online Service

Product Details

High Resistance / Low Conductance Meter HRLC Description:

There is always a problem with ultra high resistance measurements: only a very small amount of current is flowing through the component that is being tested. Therefore certain precautions must be taken to prevent interference from external sources. A grounded conductive box with a copper mesh viewing screen allows testing of small components without interference from stray static electricity (usually caused by the operator’s movements). This box, which attaches to the meter, is included. In addition, the “sensitive” terminal (the terminal which is sensitive to static electric fields) can be connected through a shielded test cable (included) to measure any assembly that can’t fit in the conductive box.

The meter’s six resistance ranges are 19.999/ 199.99/1999.9 MegOhms and 19.999/199.99/1999.9 GigOhms. The highest direct resistance that can be read is 1999.9 GigOhms, and the minimum resolution is .001 MegOhm (1 KiloOhm), when the meter is set for reading 19.999 MegOhms. Overall accuracy is +/-2% of the reading, +/- one count.

In addition, there are two conductance (inverse of resistance) ranges. They are 19.999 NanoSiemens and 19.999 PicoSiemens. A NanoSiemen is 1 divided by a GigOhm. It’s the same as one NanoAmp per Volt. A PicoSiemen is one PicoAmp per Volt, or the inverse of one TeraOhm. Therefore, a 1 GigOhm resistor will have a conductance of 1 NanoSiemen, and a 1 TeraOhm will correspond to 1 PicoSiemen. Note that because one is the inverse of the other, then 2 TeraOhms corresponds to 0.5 PicoSiemen. The two conductance ranges therefore cover the following ranges of equivalent resistance: The first conductance range goes from .001 x 10-9iemens (=1010 Ohms or 1 TeraOhm) to 19.999 x 10-9 Siemens (= about 5 x 106 Ohms or 5 MegOhms), and the second conductance range goes from .001 x 10-12 Siemens (=1015 Ohms or 1000 TeraOhms) to 19.999 x 10-12 Siemens (= about 5 x 10-9 Ohms or 5 GigOhms). Accuracy is +/-2% of the reading, +/- one count, but external interference may cause additional inaccuracy.

It’s easier to measure very high resistances by using the conductance method than by using the standard resistance measurement method. With a standard resistance measurement, a certain pre-set amount of current is passed through the resistor that is being measured. For example, when the meter is set for 19.999 MegOhms, the internal circuitry causes exactly 100 NanoAmps (10-7 Amps) to flow through the resistor. A 10 MegOhm resistor would then have a voltage of 1 Volt (=107Ohms x 10-7 Amps) across it, and the display will read “10.000” as a result. When the meter is set to any of the 6 resistance ranges, the display reads proportional to whatever the voltage is across the test resistor. One volt always produces a half-full-scale reading, and 1.999 Volts always produces a full-scale reading of 19999, but the decimal point position depends on which scale is being used. There are 6 different pre-set amounts of current used: Starting with the 19.999 MegOhm and ending with the 1999.9 GigOhm range, the pre-set currents are 100,10, and 1 NanoAmp; and 100,10, and 1 PicoAmp. In each case, if the resistor voltage happens to be exactly 1.9999 Volts (when the appropriate amount of current is passing through it) the meter will show exactly full scale. When measuring conductance, a different technique is used: A voltage difference is applied across the resistor and then the current flowing through the resistor is measured. For purpose of illustration, think of the voltage as 1 Volt. Then the conductance (in Siemens) is the same number as the current (measured in Amps). For example, if a 10 GigOhm (1010 Ohms) resistor has 1 Volt applied across it, a current of 0.1 NanoAmp (10-10 Amp) will flow through the resistor, and this means its conductance is 10-10 Siemens or 0.1 NanoSiemen.

There are two major differences between resistance and conductance measurements:

1) A resistance measurement may take a very long time to settle to its final value if there is very much capacitance connected across the resistor;

2) By the nature of an Ohm reading vs. 1/Ohms, if a very high resistance is being measured (a resistance which measures near full-scale on the meter), and the resistance is fluctuating by a factor of 10% from one second to the next (it may be noisy because a high resistance has a very “weak” signal), then the display will fluctuate 10% of 19999 counts, or about 2000 counts. This is very difficult to read. A low resistance (which is usually much less noisy) would show a very low number on an Ohm setting, so it would be much more stable. This means that when doing resistance measurements, small resistances are extremely stable, while large resistances are disproportionately unstable. However, if the meter is switched to conductance measurement, a high resistance will read a low number on conductance, and a 10% fluctuation may not even be visible as a one-count fluctuation.

Here’s an example of how capacitance can slow down the resistance reading: when measuring a 1 TeraOhm resistor, if less than 1 PicoFarad of capacitance is present, the resistance measurement will settle to within 2% of the final value in 4 seconds, and conductance measurement will settle in 3 seconds. However, if 10 PicoFarads is connected in parallel with the 1 TeraOhm, then resistance measurement requires 40 seconds to settle, while conductance still requires only 3 seconds. (Please note that these settling times apply to air-gap capacitance in the circuit. Solid capacitors often require a longer time to come to equilibrium, because they polarize slowly.) The slow settling of resistance measurements is because the tiny amount of current passed through the capacitor-resistor combination requires a long time to charge the capacitor up to its final voltage.

There are additional control knobs on the meter. The RESPONSE SPEED (signal averaging time) can be set to FAST (recommended) or SLOW (if the reading on FAST fluctuates too much). The previous quoted settling times were on FAST.

There are two offset controls:

1) a zero-ohms offset (accessed through a small hole on the right side of the meter), which is adjusted to make the display read zero when the two test terminals are shorted. This zero can be done on any of the 6 resistance ranges and it does not need readjustment when switching to another resistance range. Readjustment is only needed if the temperature changes more than 30o F.

2) A zero-conductance offset (larger knob on right edge). This is adjusted to read zero when the terminals are not connected, and only when the RANGE knob is set on [CONDUCTANCE]. It must be readjusted when switching the toggle switch between [NanoSiemens] and [PicoSiemens]. This subtracts out the tiny currents (usually a few FemtoAmps) from the amplifier’s input, and from other mechanical effects that produce weak currents. This offset should be adjusted only after the other (small knob) offset is done, and should be adjusted more frequently than the small knob offset is adjusted.

Instructions:

? Calibration of zero-ohms

The zero-ohms offset knob must first be set correctly. Plug the short alligator clips into the two terminals. The alligator clip with the larger diameter plug goes into the “Non-Sensitive Terminal”. To avoid electrostatic damage to the meter, it’s best to touch one of the two “shield” jacks before plugging in the “Sensitive Terminal”, and to do that anytime just before touching the “Sensitive Terminal”. Connect the alligator clips together by rotating them until they touch, and then clip one on the other. Turn the RANGE knob to any one of the numbers (but not to CONDUCTANCE), and the UPDATE SPEED knob to FAST (this will also turn the meter on). Note the knob sticking out of the right side of the meter. There is a hole in the case just next to that knob. Use a small screwdriver (included) to adjust the control at the bottom of the hole until the display reads zero. This adjustment will probably never need to be done again, but you should check once or twice a year to see if it really remains zero.

? Resistance measurement

Connect the unknown component (resistance) between the “Non-Sensitive” and “Sensitive” terminals, using the alligator clips. (If the component is small enough, you will not need to use the two cables; small components can be held directly over the meter by the alligator clips.) Snap the shielded box’s plug into either the left or right jack that says “shield”, making sure (by looking through the box’s screen) that none of the test assembly touches the inside of the shielded box. If the component is too big to fit inside the shielded box, use the two cables to extend the reach of the alligator clips. Note that the cable that fits the “Non-Sensitive Terminal” has only one conductor, whereas the “Sensitive Terminal” cable is coaxial (inner + outer conductors). If a long, unshielded wire were accidentally to be connected to the “Sensitive Terminal”, then the meter would fluctuate whenever the electric field in the room changes, even if that change is very small. If the electric field stops changing (either in strength or direction) for over 3 seconds, then the meter will become stable and it will read the correct amount. Because of this sensitivity to external fields, the “Sensitive Terminal” has a coaxial cable. The outer conductor shields the inner conductor. Note that anything connected (at the far end of the cable) to the terminal of this inner conductor should be shielded as much as possible from electric field in the room.

There are three separated voltage levels coming from the meter that should not be connected together, even though they all appear to be “case ground”: 1) The two banana connectors that say “shield” are true case ground. 2) The outer conductor of the Non-Sensitive Terminal” is almost at exactly the same voltage as the case ground, but an amplifier holds it near case ground voltage. The metal shaft of the “MegOhms”, “GigOhms” toggle switch is also at this potential. This is potential called the “guard ring”. 3) The “Non-Sensitive Terminal” is between zero and two Volts below case ground when measuring resistance. When measuring conductance, it is 3.5 Volts above the case ground. Never connect 1) to 2), 1) to 3), or 2) to 3).

To make a resistance measurement, switch the RANGE to 1999.9 and the toggle switch to GigOhms. If after a few seconds, the display fluctuates up and down significantly, switch the RESPONSE SPEED to SLOW. If the display instead shows a “1” on the far left side, it means the resistance is greater than 1999.9 GigOhms. (In that case, only the CONDUCTANCE will give a reading, as will be described later). However, if the display reads a number (1999.9 GigOhms or less), this is the correct resistance. If the number is very small (or zero), you may want to switch the RANGE knob from 1999.9 down to 199.99 or 19.999; if the number displayed is still small, switch the toggle switch from GigOhms down to MegOhms, and go through the 1999.9, 199.99, 19.999 ranges again until you see a large enough number to read with whatever level of resolution you need. If you go too far, a “1” will appear on the left side of the display, and no other digits will be displayed until you switch the range back again.

There may be several problems in getting a steady, precise number to be displayed.

1) If a lot of capacitance is present (which is expected if you are measuring the resistance of a thin sheet of insulator that is sandwiched between the two electrode sheets), then the measured resistance may increase very slowly instead of immediately rising to the correct value. This problem with the resistance slowly rising (and taking a long time to become stable) is most likely to occur when on the highest resistance range: 1999.9 GigOhms. If the time required is unacceptably long, the CONDUCTANCE range will need to be used.

2) Another effect on stability of the displayed number may arise from external DC electric fields that are changing, while the test assembly is not properly shielded from these fields. ‘Proper shielding’ means that the assembly is surrounded by metal foil or metal screen which is itself connected to one of the meter’s two “shield” connectors.

3) A third problem has to do with the components that are being tested. These may retain charge in certain regions, and these regions are usually slow to discharge. An example of this that was mentioned earlier is a real-world capacitor. For example, a 10 PicoFarad real-world capacitor charged to 1 Volt may be discharged through a 1 TeraOhm capacitor. This combination of resistor and capacitor should discharge to ? of its previous value every 7 seconds. That is, after 7 seconds, the capacitor should retain ? Volt; after 14 seconds, it should retain ? Volt, etc. At first, a real-world capacitor will act this way – cutting its voltage in half every 7 seconds. By the time the voltage is down to around 1/100 of a Volt, it may slow down, and cut its voltage in half every 15 seconds. Eventually it gets slower and slower, so that the last bit of charge takes an unexpectedly long time to remove. This will cause the meter display to take a very long time to settle to its final value. There may also be charged insulating surfaces on the components that are being tested. The slow discharge of these surfaces, even if the surfaces are not directly connected to the “Sensitive Terminal”, will cause a changing external field, which causes the display to read either too high or too low for a long interval of time before it finally settles. If an insulating surface of a ‘component that is being tested’ is accidentally rubbed, it may acquire static charge, and this problem (above) will result. Generally, any static electricity on the component can be removed before measurement is done, by using a neutral ionizer or by dipping the component in a (grounded) container of water and then shaking dry. The first effect, 1) (mentioned above), having to do with capacitance, can be overcome by using CONDUCTANCE. The second effect, 2) (also mentioned above), having to do with unexpectedly slow discharge of capacitance, in which the last little bit of charge is much slower to remove than it should be, cannot be overcome by using either resistance or conductance. If this second effect is a problem, try to reduce any ‘capacitance that involves solid insulators’ from the assembly you are measuring. Air-gap capacitance will not cause this second effect, so air-gap capacitance need not be eliminated. Another problem arises from measuring the resistance of ordinary materials that are not usually used to make resistors. With many types of material, the resistance depends on how much voltage is applied across the material. That is, a given sample may have 1 Volt across it when 1 NanoAmp is passing through it; but 2 Volts when 3 NanoAmps are passing through it. However, in an ideal resistor, if 1 Volt corresponds to 1 NanoAmp, then 2 Volts would correspond to 2 NanoAmps. Many materials will have a lower and lower resistance as you raise the voltage across the terminals.

? Conductance Measurement

If the resistance is higher than 1999.9 GigOhms, or if the capacitance of the circuit is so high that the resistance reading is drifting up too slowly to wait for it to stabilize, as in the first effect 1), use the conductance method of measurement. Switch the RANGE knob to [CONDUCTANCE]. There are two ranges of conductance: 19.999 NanoSiemens and 19.999 PicoSiemens. A Siemen is a measure of the amount of current (in Amps) that will flow through the component if 1 Volt is applied across it. A component with a conductance of .4 PicoSiemens means that .4 PicoAmp will flow through it if 1 Volt is applied across it. This component may therefore be called either one of the following: a 2.5 TeraOhm resistor or a .4 PicoSiemen conductor.

Before taking a reading, the ‘conductance offset’ must be adjusted correctly. This offset must be adjusted more frequently than the ‘zero-ohms offset’ that was previously adjusted. ‘Conductance offset’ is controlled by the knob sticking out of the right side of the meter. A circuit which has “zero conductance” is an open circuit. (This is in contrast to “zero resistance”, which is a short circuit.) Therefore, to zero the conductance, you must first disconnect the two test terminals (Sensitive and Non-Sensitive) from each other, and of course also disconnect (on at least one side) any component being tested. Leave the alligator clip (or coaxial cable) plugged into the “Sensitive Terminal”, but disconnect everything else from that Sensitive Terminal’s clip. If there is a component that is ready to test, one of its leads can remain connected to the “Non-Sensitive Terminal”. This will minimize the number of movements needed to connect the component for testing after a correct ‘zero conductance offset’ is done. Avoid touching or rubbing the plastic ring around the “Non-Sensitive Terminal”, or the plastic ring separating the inner from outer conductor on the “Sensitive Terminal”. Static charge could build up and it would require some time to dissipate.

Once the component is ready to test, but not yet connected to the “Sensitive Terminal”, plug the conductive box into its place, if possible. Switch the toggle switch to [PicoSiemens] (same position as GigOhms). Then rotate the conductance offset knob (sticking out of right face of meter) until the display reads near zero. This is the correct setting when the toggle is on [PicoSiemens], but the knob may need a different setting if the toggle is later set to [NanoSiemens] (re-zero the conductance offset whenever you change the toggle switch setting). Then remove the conductive box (if it’s used) and connect the component to test. Remember that in this conductance setting, a small number means high resistance, and a large number means low resistance. The conductance offset may shift by one count or more if the meter’s temperature changes by at least 2 degrees F. It may also shift by several counts if there is some charge on the surfaces of the test assembly; as this charge dissipates, the offset level shifts slightly. In the most sensitive toggle position (PicoSiemens), each Count represents a change of 1 FemtoAmp flowing through the component if 1 Volt is applied across that component. (In reality, 3.5 Volts is across the component, so each count on the display actually represents a current differential of 3.5 FemtoAmps.) Because a FemtoAmp (10-15 Amp) is so small, the assembly is very susceptible to external electric fields and to temperature changes which will affect the amplifier’s offset current. The conductive box should be used if possible. Again, avoid touching or rubbing any insulating surfaces when connecting the component to be measured. If the reading is initially unstable, check the shielding and/or stand very still to avoid movement in the room’s electric field.

High Resistance / Low Conductance Meter HRLC Specification:

Measures resistance in 6 ranges from 19.999 MegOhms to 1999.9 GigOhm Accuracy is +/-2% of the reading +/- 1 count The 6 measurement currents are 100 NanoAmps down to 1 PicoAmp in factors of 10
Measures conductance in 2 ranges: 19.999 NanoSiemens and 19.999 PicoSiemens Accuracy is +/-2% of reading +/- 1 count During conductance measurements, 3.5 Volts is applied across the component
A “RESPONSE SPEED” control allows selection of either FAST meter update (1/3 second time constant), or SLOW (2 second)
Noise: for resistance greater than 10 GigOhms, RMS noise per 1/3 sec sample (FAST) is 0.0007 PicoSiemens. For less than 10 GigOhms, RMS noise per 1/3 sec sample is (10 GigOhms/R)1/2 x 0.0007 PicoSiemens The noise per 2 sec sample (SLOW) is approximately half as much The effect of this thermal noise on the displayed resistance can be calculated by first converting the expected resistance value to conductance (by inverting). Then add or subtract the conductance noise and convert back to resistance, These resulting “added” and “subtracted” numbers are the 70% confidence limits of the displayed resistances, ignoring outside interference
Battery is a standard rectangular 9 Volt type, with “LOW BATTERY” appearing on the display when approximately one hour of battery life remains Battery life is about 40 hours using an alkaline, and 25 hours using a regular battery A battery is included
Warranty for this meter is one year. Made by AlphaLab,Inc. (USA)

WeChat

Customer Service QQ

Customer Hotline:

0l0-52867771
l3811111452
l7896OO5796

Technical Supports

l7896OO5796
info@dorgean.com
色婷婷av国产精品 | 98精品国产自产在线观看 | www.狠狠操 | 国产高清无av久久 | 一区二区理论片 | 午夜精品久久久久久久99 | 色夜视频 | 香蕉色综合 | 国产一区av在线 | 亚洲精品理论片 | 免费在线观看视频一区 | 精品免费视频 | 久久久免费看 | 天天干天天干天天色 | 91九色蝌蚪视频在线 | 久久大片 | 国产视频一区二区在线观看 | 亚洲精品久久久蜜臀下载官网 | 久久大片 | 婷婷国产v亚洲v欧美久久 | 久久中文字幕在线视频 | 国产精品久久久久三级 | 十八岁免进欧美 | 91九色在线视频 | 精品中文字幕在线播放 | 日本黄色免费网站 | 国内精品福利视频 | 91日韩在线| 有没有在线观看av | 国产精华国产精品 | 色婷婷激情电影 | 在线观看亚洲免费视频 | 99re8这里有精品热视频免费 | 人人搞人人爽 | 999久久久久 | 丁香视频五月 | 成人黄色av免费在线观看 | av久久在线 | 亚洲日本va中文字幕 | 欧美精品在线免费 | 久久999精品 | 久视频在线| 婷婷激情五月综合 | 国产精品理论片在线播放 | 日韩精品视频在线观看免费 | 久久精品精品电影网 | 99久久久久成人国产免费 | 亚州国产精品久久久 | 在线观看视频国产一区 | 日韩在线观看影院 | 国产精品久久久久9999 | 免费观看www视频 | 麻豆一精品传二传媒短视频 | 激情网色 | 日韩av看片| 亚洲国产最新 | 国产精品18videosex性欧美 | 成人久久久久久久久久 | 亚洲最大成人免费网站 | 国产精品嫩草69影院 | 激情视频免费在线观看 | 99久久精品国产欧美主题曲 | 麻豆va一区二区三区久久浪 | 国产日韩中文字幕在线 | 日p视频 | 久久精品一二三 | 日日干综合 | 国产破处在线播放 | 日韩在线一二三区 | 九七人人干 | a级片韩国 | 国产免费不卡 | 国内成人精品2018免费看 | 婷婷综合亚洲 | 91成版人在线观看入口 | 免费视频三区 | 婷婷丁香色综合狠狠色 | 成人网在线免费视频 | 日韩大片在线观看 | 欧洲成人av| 一级片免费观看 | 中文字幕精品一区 | av一区二区在线观看中文字幕 | 亚洲欧美少妇 | 国产色拍拍拍拍在线精品 | 免费看搞黄视频网站 | 成人午夜电影网站 | av免费电影网站 | 国产成人精品午夜在线播放 | 亚洲综合日韩在线 | 亚洲精品一区中文字幕乱码 | 四虎成人在线 | 国产99久久久国产精品成人免费 | 精品欧美乱码久久久久久 | 在线观看亚洲专区 | 日日操夜夜操狠狠操 | 国产亚洲精品综合一区91 | 天天做日日做天天爽视频免费 | 亚洲成免费 | 国产精品久久久久999 | 视频在线一区 | 97在线精品国自产拍中文 | 久久久久久久国产精品 | 中国一级片视频 | 伊人五月婷 | 色精品视频 | 国产精品ssss在线亚洲 | 国产精品日韩欧美一区二区 | .国产精品成人自产拍在线观看6 | 国产精品夜夜夜一区二区三区尤 | 亚洲蜜桃在线 | 亚洲 综合 国产 精品 | 日韩午夜小视频 | 亚洲午夜剧场 | 国产在线欧美 | 五月婷久 | 色欧美88888久久久久久影院 | 99久久这里有精品 | 免费观看的黄色 | 玖玖在线免费视频 | 丁香婷婷久久久综合精品国产 | 欧美日韩一区二区三区免费视频 | 国产精品麻豆免费版 | 超碰97中文 | 国产精品成人在线观看 | 国产精品久久久久免费a∨ 欧美一级性生活片 | 91福利视频久久久久 | 久久久精品 | 在线不卡a | 亚洲激情电影在线 | 免费av观看| 日韩在线视频精品 | 国产字幕在线看 | 国产精品99久久久久的智能播放 | 亚州精品成人 | 亚洲综合五月 | 欧美精品黑人性xxxx | 91中文字幕在线观看 | 毛片网站免费 | 精品一二三四视频 | 狠狠色丁婷婷日日 | 亚洲综合网站在线观看 | 欧美一区在线观看视频 | 97综合网 | 一个色综合网站 | 精品国产欧美一区二区三区不卡 | 黄色网www| 色综合a | 人人添人人 | 久一在线| 天天弄天天干 | 亚洲精品日韩av | 日本精a在线观看 | 麻豆精品在线 | 免费一级特黄毛大片 | 日韩成人免费电影 | 成人av免费电影 | 久久五月天色综合 | 午夜av免费在线观看 | 久久久久久久久免费视频 | 久久看视频 | 久久国产精品偷 | 激情网综合 | 国产精品久久久久久久久久三级 | 久久精品香蕉视频 | 欧美日韩有码 | 久久精品aaa | 国产精品久久久久免费观看 | 人人爽人人乐 | 在线观看免费福利 | 91福利视频网站 | 亚洲视频网站在线观看 | 韩国av免费观看 | 91探花在线视频 | 成 人 黄 色 视频免费播放 | 国产精品黄色 | 亚洲精品久久久久久久不卡四虎 | 中文av在线播放 | 亚洲乱码久久久 | 亚洲综合在线五月天 | 中文字幕超清在线免费 | 欧美黄色软件 | 丰满少妇在线观看网站 | 久久久久久久av麻豆果冻 | 五月天狠狠操 | 国产一区在线观看免费 | 国产精品免费不卡 | 九九久久免费视频 | 国产精品综合久久久 | 又色又爽又激情的59视频 | 欧美一区中文字幕 | 久久精品视频一 | 国产成人一区二 | 91看片淫黄大片一级在线观看 | 国产精品理论片在线播放 | 五月天电影免费在线观看一区 | 2021久久 | 国产色视频一区 | 激情五月播播久久久精品 | 在线免费观看视频一区 | 激情综合亚洲精品 | 色婷婷久久 | 国产一区二区久久久 | 亚洲精品午夜aaa久久久 | 在线看黄色av | 91爱在线 | 91看片在线免费观看 | 久久精品国产一区二区电影 | www久久国产 | 国产中文欧美日韩在线 | 在线看片a | 人交video另类hd | 国产一区二区三区 在线 | 亚洲成人av一区二区 | 在线观看国产日韩 | 日韩中文字幕一区 | 国产精品成久久久久三级 | 亚洲第一中文字幕 | 免费看片网址 | 亚洲国产成人精品电影在线观看 | 婷婷久草| 久章草在线 | 久操视频在线免费看 | 国产91综合一区在线观看 | 最新日韩视频在线观看 | 国产精品原创视频 | 日韩精品资源 | 色狠狠操 | 国产一区麻豆 | 在线看91| 日韩欧美精品在线 | 国产精品丝袜久久久久久久不卡 | 99精品在线观看 | 天天天天射 | 99国产精品一区 | 欧美精品亚洲二区 | 国产精品久久久久久久久久免费看 | 国产精品人人做人人爽人人添 | 久色伊人| 精品一区二区久久久久久久网站 | 亚洲欧洲一级 | 国产一区国产二区在线观看 | 青青河边草手机免费 | 在线观看色网站 | 99视频黄 | 久久99精品久久久久久秒播蜜臀 | 在线午夜 | 欧美午夜久久久 | 欧美激情亚洲综合 | 久久久久久久久久久久久9999 | 精品一区三区 | 99精品电影 | 亚洲精品国产精品国自产 | 在线播放日韩 | 欧美日韩伦理在线 | 国产免费成人av | 色先锋资源网 | 欧美一区二区三区四区夜夜大片 | 五月天久久综合网 | 看毛片网站 | 四虎精品成人免费网站 | 精品国产电影一区 | 四虎在线观看精品视频 | 九九热在线视频 | 日韩在线播放视频 | 在线免费三级 | 亚洲三级av | 日韩理论在线播放 | 91成人精品在线 | 五月激情在线 | 国产视频色 | 日韩日韩日韩日韩 | 免费91麻豆精品国产自产在线观看 | 国产一区黄色 | 国产成人三级三级三级97 | 中国老女人日b | 国产日韩在线看 | 国产视频一区二区在线播放 | 亚洲精品女人 | 久久99九九99精品 | 成人cosplay福利网站 | 精品国模一区二区 | 成年美女黄网站色大片免费看 | 五月天婷亚洲天综合网鲁鲁鲁 | 亚洲资源在线 | www.天天射 | 色婷婷国产在线 | 我要色综合天天 | 99热在线免费观看 | 久久精品一二三区白丝高潮 | 国产精品欧美 | 国产成人精品a | 久久福利小视频 | 黄色a一级片 | 色在线网 | 97超碰人人看 | 麻豆系列在线观看 | 人人插人人艹 | 99在线观看视频网站 | 在线观看亚洲专区 | 国产日女人 | 夜色在线资源 | a在线观看免费视频 | 亚洲日本中文字幕在线观看 | 色噜噜在线观看 | 国产成人精品久久亚洲高清不卡 | 亚洲精品玖玖玖av在线看 | 最新中文字幕在线资源 | 亚洲一级片在线看 | 精品国产美女在线 | 国产日韩精品欧美 | 综合网伊人 | 在线观看免费av网 | 日韩欧三级 | 日韩免费观看一区二区 | 国产91精品一区二区麻豆亚洲 | 丁香综合五月 | 精品久久久久一区二区国产 | 黄色一级性片 | 日韩欧美一区二区三区视频 | 91在线影院 | 欧洲精品亚洲精品 | 国内精品免费 | 日日草av | 国产在线观看中文字幕 | 免费观看性生活大片 | 伊人资源站 | 美女网站色在线观看 | 久久久亚洲成人 | 天天射天天艹 | 182午夜在线观看 | 欧美激情一区不卡 | 91中文字幕在线 | 久草在线视频免费资源观看 | www久久精品 | 天天色婷婷 | 91麻豆精品国产91久久久更新时间 | 欧美一进一出抽搐大尺度视频 | 欧美伦理电影一区二区 | 亚洲激情 在线 | 91亚洲精品久久久蜜桃 | 亚洲精品资源 | 国产探花 | 免费 在线 中文 日本 |